Description
- 1/4 DIN watertight, IP65 rated enclosure
- Reversible & isolated 4 to 20 mA output with adjustable bandwidth
- Power down, non-volatile memory backup
- Multiple spot on/off relays
- 1 to 5 point calibration
- Graphical LCD display with backlight
- Electrode self-cleaning function
- Embedded self diagnostic program
- RS-485 Modbus output
- 4-digit password security protection
- High 50/60 Hz noise rejection
- CE compliant
Specifications
Turbidity Range 1
0.00 to 100.00 NTU (FTU) (TC-100)
Turbidity Resolution 1
0.01
Turbidity Accuracy 1
± 3% F.S.
Turbidity Self Cleaning 1
30 Mins.
Turbidity Range 2
0.0 to 500.0 NTU (FTU) (TC-500)
Turbidity Resolution 2
0.1
Turbidity Accuracy 2
±3% F.S.
Turbidity Self Cleaning 2
10 Mins.
Turbidity Range 3
0 to 3000 NTU (FTU) (TC-3000)
Turbidity Resolution 3
1
Turbidity Accuracy 3
±3% F.S.
Turbidity Self Cleaning 3
10 Mins.
Suspend Solid Range 1
0.0 to 1000 mg/L (TCS-1000)
Suspend Solid Resolution 1
1
Suspend Solid Accuracy 1
±3% F.S.
Turbidity Self Cleaning 1
10 Mins.
Suspend Solid Range 2
0 to 50000 mg/L (TCS-MXS-A)
Suspend Solid Resolution 2
1
Suspend Solid Accuracy 2
±3% F.S.
Turbidity Self Cleaning 2
10 Mins.
Repeatablity
±2% F.S.
Calibration
1 to 5 point
Current Output
One set of isolated and reversible 4 to 20 mAcurrent output. Max load 500 Ω
Digital Output
Isolated RS-485 modbus output
ON/OFF Relays
Two ON/OFF relays for turbidity;one ON/OFF for electrode self cleaning 5 A at 115 VAC or 2.5 A at 220 VAC
External Cleaning Controller
ON 0 to 9999 seconds;OFF 0 to 999.9 hours. Self programmable
Electrde Self Cleaning
10 or 30 minutes
Power
110 or 230 V AC, 50/60 Hz
Dimensions
1/4 DIN, 96 x 96 x 148 mm